The new X-Mind AC generator is fitted with a Toshiba DG-073B monophased auto rectifier tube. It generates a high 70kV voltage with an ideal intensity of 8mA. Concentrated, high density radiation in a very small diameter focal spot (0.7mm) for greater accuracy. Supplied with a long cone as standard, the generator is perfectly suited to the parallel technique (a short cone and a rectangular cone are available as options). The rays emitted have a large wavelength and provide pictures characterized by a widespread grey scale for better distinction of anatomical detail. The new timer is programmed for use with D, E or F type films. It can be installed remotely and can control one or two units whatever the technology (AC + AC or AC + DC), making the X-Mind concept unique and providing the most flexibility of use in intra-oral radiography.
Classification : Electromedical equipment, Class 1 type B
Supply voltage : 220/230/240V - monophase 50/60Hz
Power absorption at 230V : 0.8kVA
Resistance of the line : 0.5 ohms
X-ray tube : New Toshiba DG 073B tube
High voltage : 70kV
Anode current : 8mA
Focal spot : 0.7mm
Total filtration : Equivalent to 2mm Al at 70kV
Tube inherent filtration : Equivalent to 0.8mm Al at 70kV
Leakage radiation : Less than 0.25mGy/hour maximum at a distance of 1m from focal spot
Long cone : Focus to cylinder tip distance = 31cm (12)
Timer : Exposure time can be set from 0.08 to 3.2 seconds
X-ray emission control : Dead man switch with 3m spiral cable
Total weight : 28kg (weight of the head : 9kg)